Err# Message Verbose/Correction Information
-300 Device-specific error An instrument error occurred and the exact problem cannot be
specifically identified. Report this error to the nearest Agilent
Technologies sales or service office.
-310 System error; An internal system-type error has occurred. The exact problem
cannot be specifically identified. Report this error to the
nearest Agilent Technologies sales or service office.
-310 System error; A license
will soon
expire;<feature code>
will expire in <time>
The indicated feature/software will expire in the specified
time. Contact Agilent Technologies to purchase continued use
of this functionality.
-310 System error; Error
transmitting a LAN event
to the network.
Communication with the network driver failed.
-310 System Error; Ext Source
not responding
Signal source at given IP address is not responding / IP does
not belong to a source. Check IP address and network
connection.
-310 System error; Failed to
initialize the PTP clock
to current time.
Failure communicating with the DMC libraries’ PTP
controller.
-310 System error; Failed to
instantiate the PTP
ordinary clock.
Failure in the starting up the DMC libraries’ PTP controller.
-310 System error; feature
<feature code> not
licensed
The specified feature, for example “N9073A-TR2” is not
licensed. The license may have expired. You cannot use it until
you get a license.
-310 System error; Feature
expired; <feature code>
The specified feature has expired. The license is no longer
valid.
-310 System error; License
installation
failed;<feature code>
The license installation of the specified feature, for example
"N9073A-TR2", has failed. You should refer to the event log in
the control panel for more details.
-310 System error; License
removal failed; <feature
code>
The license removal of the specified feature, for example
“N9073A-TR2” has failed. You should refer to the event log in
the control panel for more details.
-310 System error; No
license; <feature code>
will terminate in <time>
The specified feature will stop working in the specified time
due to the license expiration You will be prompted to save