242 Chapter 13
Assembly Replacement Procedures
Power Supply Assembly
Replacement
1. Slide the power supply assembly into the slot at the rear of the instrument and
push on the assembly to mate the connectors to the midplane assembly.
2. Refer to Figure 13-10. Replace the three screws (1) through the power supply
bracket (3) and into the power supply (2). Torque to 9 inch-pounds.
3. Replace the CPU assembly. Refer to the CPU Assembly replacement
procedure.
4. Replace the reference assembly. Refer to the L.O. Assembly replacement
procedure.
5. Replace the top brace and reference bracket. Refer to the Top Brace and
Reference Bracket replacement procedure.
6. Replace the rear panel. Refer to the Rear Panel replacement procedure.
7. Replace the instrument outer case. Refer to the Instrument Outer Case
replacement procedure.