Chapter 15 299
Functional Tests
Second Harmonic Distortion (SHD)
Figure 15-3 Second Harmonic Distortion Test Setup
Procedure
1. Configure the equipment as shown in Figure 15-3.
2. Press
Mode, Spectrum Analyzer, Mode Preset on the signal analyzer and
Preset the synthesized sweeper.
3. Set up the synthesized sweeper by pressing:
Frequency, 40, MHz
Amplitude
, –10, dBm
4. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel
, Center Freq, 40, MHz
SPAN X Scale
, Span, 1, MHz
5. On the analyzer, press Peak Search.
6. Adjust the synthesized sweeper amplitude for a signal analyzer display of
–10 dBm ±0.1 dB.
7. On the analyzer, activate the marker delta function by pressing the
Marker and
Delta keys.
8. On the signal analyzer, press:
FREQ Channel, Center Freq, 80, MHz
Meas Setup
, Average/Hold Number, 20, Enter
Trace/Detector, Trace Average
Single
9. Press Peak Search. Enter the displayed value under the Measured Second
Harmonic Distortion (dBc) heading in Table 15-3.
Table 15-3 Second Harmonic Distortion Results
Applied Frequency
(MHz)
Measured Second Harmonic Distortion
(dBc)
Mixer Level
(dBm)
Specification
(dBc)
40 –20 –55