94 Chapter 3
Instrument Messages
-200 to -299, Execution Errors
-200 to -299, Execution Errors
For -221 error messages, see the previous sections.
Note that Execution Errors are divided into subclasses:
-21x – Trigger errors
-22x – Parameter error
-23x – Data corrupt or stale (invalid data)
-24x – Hardware error
-25x – Mass storage error
-26x – Expression data error
-27x – Macro error
-28x – Program error (a downloaded program-related execution error)
-29x – Memory use error
Err# Message Verbose/Correction Information
-200 Execution Error A program execution error has occurred. The exact problem
cannot be specifically identified.
-200 Execution error; Carrier
frequency outside
device’s transmit band
The entered channel/carrier frequency is not within the range of
your current mode setup selection of standard and device.
-200 Execution error; Invalid
GSM burst timing
A GSM-like burst was acquired, but its timing is not valid.
Ensure the correct Burst Type has been selected.
-200 Execution error; Invalid
IP address
The IP address supplied is either not valid or does not belong to a
compatible Signal Generator. Please check the IP address and
instrument connection and try again.
-200 Execution error; Invalid
Marker Trace.
Cannot place markers on the reference trace, because the
reference trace is currently turned off or has no data.
-200 Execution Error; No peak
found.
No signal peak was found within the defined parameters of the
search.
-200 Execution error; No ranges
are defined. Activate a
range.
There are no active ranges in the range table. You will need to
activate at least one range.
-200 Execution Error;
Preselector centering
failed
Algorithm failed to center the preselector. This maybe caused by
the signal peak being too low in amplitude. Or it could be from
excessive CW input signal, alignment error, or hardware failure.
-200 Execution Error; Signal
not stable enough to track
The signal that you have selected to track is changing too much
for the function to track it properly.