Instrument Messages
-200 to -299, Execution Errors
-200 Execution Error; Store ref
trace before turning on
Normalize
The Reference trace data must be stored in the Ref trace before
you turn on the Normalization function.
-200 Execution error; Sync word
was not found.
NADC & PDC: In an EVM measurement, the sync word is not
found and the synchronization cannot be established when Sync
Word is selected in the Burst Sync menu.
Flexible Digital Demodulation: The sync word cannot be
detected because of inappropriate parameter settings or incorrect
signal.
-200 Execution error; Trace
file contains no
compatible traces.
The trace file may have been created by another version of the
Phase Noise personality, which uses a different trace format that
is incompatible with the version you are running. Please check
you are running the most up to date version of the personality.
-200 Execution error; Trace
file created by
incompatible version of
Phase Noise App
The trace file may have been created by another version of the
Phase Noise personality, which uses a different trace format that
is incompatible with the version you are running. Please check
you are running the most up to date version of the personality.
-201 Invalid while in local The command cannot be executed while the instrument in Local
control.
-202 Settings lost due to rtl A “return to local” control was forced and some settings were
lost as a result of this.
-203 Command protected The command could not be executed because it is disabled. It
was disabled by licensing or password protection
-203 Command protected;
feature not licensed
The specified feature, for example “N9073A-TR2” is not
licensed. The license may have expired. You cannot use it until
you get a license.
-210 Trigger error A trigger error has occurred, but the exact problem cannot be
specifically identified.
-211 Trigger ignored A GET, *TRG or other triggering signal was received, but it was
ignored because of timing considerations. For example, maybe
the instrument was not ready to respond when the command was
received.
-212 Arm ignored An arming signal was received, but it was ignored.
-213 Init ignored An initiate trigger/sweep request was received and ignored,
because another measurement was already in progress.
-214 Trigger deadlock The trigger source for the initiation of a measurement is set to
GET, and the following measurement query was received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
-215 Arm deadlock The arm source for the initiation of a measurement is set to GET
and the following measurement query is received. The
measurement cannot be started until a GET is received and the
GET would cause an INTERRUPTED error.