Chapter 13 241
Assembly Replacement Procedures
Power Supply Assembly
Power Supply Assembly
Removal
1. Remove the instrument outer case. Refer to the Instrument Outer Case removal
procedure.
2. Remove the rear panel. Refer to the Rear Panel removal procedure.
3. Remove the top brace and reference bracket. Refer to the Top Brace and
Reference Bracket removal procedure.
4. Remove the reference assembly. Refer to the L.O. Assembly removal
procedure.
5. Remove the CPU assembly. Refer to the CPU Assembly removal procedure.
6. Refer to Figure 13-10. Remove the three screws (1) attaching the power
supply (2) to the power supply bracket (3).
7. The power supply assembly can be removed from the chassis by pulling
straight out the back.
Figure 13-10 Power Supply Assembly Removal
1
3
2