116 Chapter 3
Instrument Messages
Condition Errors 301 to 399, Uncalibrated Integrity
Condition Errors 301 to 399, Uncalibrated Integrity
Errors with measurement calibration/alignment routines and signals.
An event with the error number shown in the table means the condition has been
detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 307 indicates an “AC: unspec’d below 10 MHz” condition has
been detected; error 1307 indicates that failure has been cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to SCPI.
Nothing in this column means nothing is put up; status bit only.
Err# Message Error or
Warning
Verbose/Correction Information
301 Meas Uncal W
303 Sig ID On W
305 No Long Code Phase W
307 AC coupled: Accy
unspec’d <10 MHz
W AC input coupling will function at lower frequencies, but
the performance is not specified below 10 MHz.
309 User cal W
309 User Cal; Adjusted for
new RBW
W The measurement RBW has been changed since the last
calibration (~CAL)