3 Spectrum Analyzer Mode
3.6 Power Stat CCDF Measurement
Name Unit Corresponding Results Results
Item for n =
1
Explanation
0.0001% dB Power level that has 0.0001%
of the power
8 99.99 dB
0.00001% dB Power level that has 0.00001%
of the power if "Minimum
Probability" on page 1138 is
PP7 (0.00001 %)
11 99.99 dB
Peak dB Peak power 9 99.99 dB
Peak dBm Not available via remote
commands
n/a 99.99 dBm
3.6.3 Amplitude
Activates the Amplitude menu and selects Reference Level or Reference Value as
the active function, depending on the measurement.
Some features in this menu apply to multiple measurements. Some other features
apply only to specific measurements and their controls are blanked or grayed-out in
measurements that do not support the feature.
3.6.3.1 Y Scale
Contains controls that pertain to the Y axis parameters of the measurement. These
parameters control how data on the vertical (Y) axis is displayed and control
instrument settings that affect the vertical axis.
Ref Level Offset
Adds an offset value to the displayed reference level. The reference level is the
absolute amplitude represented by the top graticule line on the display.
The on/off switch turns Ref Level Offset on or off. Setting a specific value turns Ref
Level OffsetON.
For more on using offsets, see the Ref Level Offset control description for the Swept
SA measurement.
Remote Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Example
:DISP:WIND:TRAC:Y:RLEV:OFFS 12.7
:DISP:WIND:TRAC:Y:RLEV:OFFS?