7 Save/Recall/Print
7.2 Recall
7.2.11.1 Recall From File
Recalls a previously saved SCPI Recorder file. For details of the SCPI Recording
feature, see "SCPI Recorder" on page 2193.
After the file contents have been read, each of the SCPI commands or queries
present in the file at the time of recall is applied to the system. If the file is from
another instrument, or from a different model, some commands may cause
unexpected data changes as each is applied. If any commands result in errors, the
command(s) and the corresponding error(s) are displayed after playback is
completed.
Recalling a SCPI recording plays the contents of the file immediately after recall.
You can view the content of the file in the SCPI recorder dialog. If there are any
entries in the SCPI recorder, you are prompted either to keep the previously
recorded data, or let it be discarded.
If you choose to discard the data, all existing recording entries are cleared, and the
SCPI recorder is populated with the recalled data.
If you choose to keep the existing recorded data, the recalled file content is
appended to the existing recording.
NOTE
Some SCPI entries in the recorded file may require the presence of other files, if
a command in the recorded file specifies the recall of other files.
7.2.12 Mask
The Mask data type is used to import and export Mask files for measurements that
use masks, such as cellular comms and real-time measurements.
7.2.13 Sequence
These need to be brought over for the EXT and/or Sequence Analyzer when they are
available in the Touch UI
7.2.14 Waveform
Recalls waveforms into the ARB memory of an Internal Source.
When you select the Waveform tab in the Save dialog,, a hint appears saying
“Recalls files from Mass Storage to the ARB and lets you manage the ARB memory
at the same time.”