9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
9.4.6.10 Questionable Calibration Extended Needed Register
Monitors conditions that occur because a calibration or alignment is required to
guarantee accurate measurements. It summarizes them in bit 8 of the "Questionable
Calibration Register" on page 2806.
Bit Condition Operation
1 Align 9kHz-
30MHz required
EMI receiver alignment required, 9kHz-30 MHz (conducted band)
2 Align 30MHz-
1GHz required
EMI receiver alignment required, 30 MHz-1 GHz (radiated band)
4 Input Attenuation
not calibrated
The input attenuator is uncalibrated
5 Align current
frequency range
required
Alignment for current set frequency range is needed. It is suggested
to process Align Selected Freq Range for the frequency range in use
8 MPA Align
required
The Multiport Adaptor must be calibrated (EXT only)
11 Characterize
Preselector
required
Go to the System, Alignments, Advanced menu and perform a
“Characterize Preselector”
12 Characterize
Noise Floor
required
Go to the System, Alignments, Advanced menu and perform a
“Characterize Noise Floor”
Filter Registers
"Questionable Calibration Extended Needed Condition" on page 2811
"Questionable Calibration Extended Needed Enable" on page 2811
"Questionable Calibration Extended Needed Event Query" on page 2812
Spectrum Analyzer Mode User's &Programmer's Reference 2810