EasyManua.ls Logo

Keysight N9041B - Δ to Limit; Query the Signal Peaks (Remote Query Only)

Keysight N9041B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
:CALC:MARK:PEAK:MPE?
Preset 20
State Saved Saved in instrument state
Min 1
Max 500
Δ to Limit
Selects the Limit to be used for the Δ to Limit column in the Peak Table, and turns
the Δ to Limit column on and off.
When ON, this column shows the difference between each peak and the specified
Limit.
Remote
Command
:CALCulate:MARKer:PEAK:TABLe:DTLimit LLINE1 | | LLINE6
:CALCulate:MARKer:PEAK:TABLe:DTLimit?
Example
:CALC:MARK:PEAK:TABL:DTL:STAT ON
:CALC:MARK:PEAK:TABL:DTL LLINE1
Preset
LLINE1
Auto Function
Remote
Command
:CALCulate:MARKer:PEAK:TABLe:DTLimit:STATe ON | OFF
:CALCulate:MARKer:PEAK:TABLe:DTLimit:STATe?
Preset
OFF
Query the Signal Peaks (Remote Query Only)
Returns a list of peaks of the specified trace that are above Peak Threshold (if ON)
and have an excursion above Peak Excursion (if ON).
Remote
Command
:CALCulate:DATA[n]:PEAKs? <threshold>,<excursion>[,AMPLitude | FREQuency |
TIME,[ALL | GTDLine | LTDLine]]
Example
:CALC:DATA1:PEAKs? -90, 6
Response:
:1.000000000E+01, -1.804702691E+01, 1.320202000E+10, -1.831304020E+01,
1.325500000E+10, -2.069423795E+01, 1.312255000E+10, -4.752705087E+01,
1.293712000E+10, -5.737315428E+01, 1.497685000E+10, -5.898005784E+01,
1.492387000E+10, -5.968565135E+01, 1.505632000E+10, -6.822387883E+01,
1.518877000E+10, -7.021781263E+01, 1.476493000E+10, -7.404477322E+01,
1.471195000E+10
Note: The first value is the number of peaks that meets the specified limits