EasyManua.ls Logo

Keysight N9041B - 3.12.3.8 Trigger Delay List; 3.12.3.9 Phase Noise Optimization

Keysight N9041B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.12 List Sweep Measurement (Remote Only)
Remote
Command
[:SENSe]:LIST:SWEep:TIME:POINts?
Example
LIST:SWE:TIME:POIN?
3.12.3.8 Trigger Delay List
This command defines a list of analyzer trigger delay time used when making the
measurements.
The query form queries the analyzer for the values in the trigger delay list.
Remote Command
[:SENSe]:LIST:TRIGger:DELay <time>[, <time>]
[:SENSe]:LIST:TRIGger:DELay?
Example
LIST:TRIG:DEL 0.01S
Preset 0 seconds
State Saved Saved in instrument state
Range Same as the Trigger Delay range for EXTernal1 trigger.
Remote
Command
[:SENSe]:LIST:TRIGger:DELay:POINts?
Example
LIST:TRIG:DEL:POIN?
3.12.3.9 Phase Noise Optimization
This command defines a list of analyzer phase noise optimization settings at which
the measurements are made.
Upon switching to List Sweep from another measurement, phase noise optimization
is set to “Fast Tune”. The user may select whatever phase noise optimization one
desires via the phase noise optimization list parameter afterwards.
Remote
Command
[:SENSe]:LIST:FREQuency:SYNThesis 1 | 2 | 3
[:SENSe]:LIST:FREQuency:SYNThesis?
[:SENSe]:LIST:FREQuency:SYNThesis:AUTO OFF | ON | 0 | 1
[:SENSe]:LIST:FREQuency:SYNThesis:AUTO?
Example
LIST:FREQ:SYNT 2
LIST:FREQ:SYNT:AUTO OFF
LIST:FREQ:SYNT 2
Preset Fast Tuning
1940 Spectrum Analyzer Mode User's &Programmer's Reference