3 Spectrum Analyzer Mode
3.9 SEM Measurement
3.9.18.3 Reference
Lets you set the Reference Power and parameters related to the Reference Power
for SEM measurements.
Measurement Type
Accesses a menu that enables you to select one of the following measurement
reference types:
Total Pwr Ref
TPRef
Sets the reference to the total carrier power and the measured
data is shown in dBc and dBm
PSD Ref
PSDRef
Sets the reference to the mean power spectral density of the
carrier and the measured data is shown in dB and dBm/Hz
Spectrum
Peak Ref
SPRef
Sets the reference to the spectrum peak power of the carrier and
the measured data is shown in dB and dBm
Remote Command
[:SENSe]:SEMask:TYPE PSDRef | TPRef | SPRef
[:SENSe]:SEMask:TYPE?
Example
:SEM:TYPE PSDR
:SEM:TYPE?
Preset WLAN Mode: SPRef
All other Modes: TPRef
State Saved Saved in instrument state
Range Total Pwr Reference | PSD Reference | Spectrum Peak Reference
Reference Power
Toggles between Measured Power (ON) and Manual Power (OFF) for Total Power Ref,
PSD Ref, and Spectrum Peak Ref.
Remote Command
[:SENSe]:SEMask:CARRier:AUTO[:STATe] OFF | ON | 1 | 0
[:SENSe]:SEMask:CARRier:AUTO[:STATe]?
Example Set to Manual:
:SEM:CARR:AUTO OFF
Set to Measured:
:SEM:CARR:AUTO ON
:SEM:CARR:AUTO?
Notes Available for all "Measurement Type" on page 1700s
1700 Spectrum Analyzer Mode User's &Programmer's Reference