6 Input/Output
6.8 Trigger Output
Source SCPI Notes
selected as the Point Trigger under Source, the Source Point Trigger under Trig 1
Out automatically gets selected. A similar pattern is used for the other Ext Trigger
inputs; for example, when Ext Trigger 2 is selected as the Point Trigger under
Source, the Source Point Trigger under Trig 2 Out automatically gets selected
Source Marker
1
S1Marker
Only available in VXT and M941xE. For M9420A, only for TRIG2, for
M9410A/11A/15A/16A available for both TRIG1 and TRIG2
Selects the Trigger Output at Marker 1 in the Waveform file that is currently playing
Source Marker
2
S2Marker
Only available in VXT and M941xE. For M9420A, only for TRIG2, for
M9410A/11A/15A/16A and M9410E/11E/15E/16E available for both TRIG1 and
TRIG2
Selects the Trigger Output at Marker 2 in the Waveform file that is currently playing
Source Marker
3
S3Marker
Only available in VXT and M941xE. For M9420A, only for TRIG2, for
M9410A/11A/15A/16A and M9410E/11E/15E/16E available for both TRIG1 and
TRIG2
Selects the Trigger Output at Marker 3 in the Waveform file that is currently playing
Source Marker
4
S4Marker
Only available in VXT and M941xE. For M9420A, only for TRIG2, for
M9410A/11A/15A/16A and M9410E/11E/15E/16E available for both TRIG1 and
TRIG2
Selects the Trigger Output at Marker 4 in the Waveform file that is currently playing
PerArb
PARB
Only available in VXT Models M9410A/11A/15A/16A and M9410E/11E/15E/16E
Selects the Trigger Output as PerArb. PerArb is a synchronization trigger which is
generated by the ARB at the beginning of each repetition of playing the signal
FSYNc
FSYNc
Only available in VXT Models M9410A/11A/15A/16A and M9410E/11E/15E/16E
Selects the Trigger Output as FSYNc, routing the Periodic Timer Sync Source signal
to the specified Trigger output. That is, the signal selected by :TRIGger
[:SEQuence]:FRAMe:SYNC is routed to the specified trigger output
The following example specifies that External 1 trigger will be used as the Periodic
Timer Sync Source, and this signal will then be routed to the Trigger 2 output:
TRIG:FRAM:SYNC EXT1
TRIG2:OUTP FSYNc
6.8.2 Trig 1 4 Out Polarity
Sets the output to the Trig n Out connector to trigger on either the positive or
negative polarity.
Remote Command
:TRIGger[1]|2|…|4[:SEQuence]:OUTPut:POLarity POSitive | NEGative
:TRIGger[1]|2|…|4[:SEQuence]:OUTPut:POLarity?
Example
:TRIG1:OUTP:POL POS