EasyManua.ls Logo

Keysight N9041B - 8.1.16 Trigger Center Frequency; 8.1.17 Trigger BW

Keysight N9041B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Trigger
8.1 Trigger
Entire Frame mode in PAL-N; if NTSC-M is selected, the line number is clipped to
525. Or, if line 313 is selected in Field 1 mode in PAL-N and NTSC-M is selected, the
line number is clipped to 263. Changing back to the PAL-N standard will leave the
line number at 263.
Remote Command
:TRIGger[:SEQuence]:TV:STANdard MNTSc | JNTSc | NTSC443 | MPAL | BPAL |
NPAL | CPAL | PAL60 | LSEC
:TRIGger[:SEQuence]:TV:STANdard?
Example Sets NTSC-M
:TRIG:TV:STAN MNTS
Queries Standard
:TRIG:TV:STAN?
Dependencies Only available in the Swept SA measurement
Only appears when TV is selected as the Trigger Source
Preset
MNTS
State Saved Saved in instrument state
Range
MNTSc|JNTSc|NTSC443|MPAL|BPAL|NPAL|CPAL|PAL60|LSEC
8.1.16 Trigger Center Frequency
Sets the center frequency to be used by the auxiliary receiver for the Auxiliary
Channel I/Q Magnitude trigger.
Remote Command
:TRIGger[:SEQuence]:AIQMag:CENTer <freq>
:TRIGger[:SEQuence]:AIQMag:CENTer?
Example
:TRIG:AIQM:CENT 10 MHz
Notes Trigger CF + 1/2 Trigger BW < Max
Trigger CF - 1/2 Trigger BW > Min
Dependencies Only appears when Aux Channel I/Q Mag is selected as the Trigger Source
Preset 0 Hz
State Saved Saved in instrument state
Range -40 MHz to 40 MHz
Min -40 MHz
Max 40 MHz
8.1.17 Trigger BW
Sets the information bandwidth used by the auxiliary receiver for the Auxiliary
Channel I/Q Magnitude trigger.
Spectrum Analyzer Mode User's &Programmer's Reference 2676

Table of Contents

Other manuals for Keysight N9041B

Related product manuals