EasyManua.ls Logo

Keysight N9041B - Minimum Acquisition Time

Keysight N9041B
2852 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Spectrum Analyzer Mode
3.9 SEM Measurement
Notes Comma separated list of values
OFFSet1 is for BTS, 2 for MS. Default is BTS
Note that Offset sub op code 2 is supported only in non-SA modes. In the SA mode, Offset sub op
code 1 is used for both BTS and MS
Dependencies On non-sweeping hardware, this column is grayed out and the Auto/Man checkbox is invisible. The
read-only value shows the estimated sweep time
In those instruments, "Minimum Acquisition Time" on page 1644 is available
Couplings When you manually set a value when in the Auto state, the state automatically changes to Man
Preset Automatically calculated
Modes other than LTEAFDD, LTEATDD, 5G NR, MSR, WLAN:
When the max number of offsets is 6:
ON,ON,ON,ON,ON,ON
When the max number of offsets is 12:
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
Modes LTEAFDD, LTEATDD, 5G NR, MSR:
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON | ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
When the max number of offsets is 14:
Mode WLAN:
ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON,ON
State Saved Saved in instrument state
Saved in instrument state
Range Auto|Man
Min Other than non-sweeping hardware
Depends on Sweep Type:
Sweep Type “Swept”: 1 ms
Sweep Type “FFT”: 100 ns
Non-sweeping hardware: N/A
Max Other than non-sweeping hardware: 4000 s
Non-sweeping hardware: N/A
Backwards
Compatibility SCPI
[:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep[:TIME]
[:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep[:TIME]:AUTO
Minimum Acquisition Time
Available on non-sweeping hardware.
Specifies the minimum acquisition time for each “chunk” of the measurement result.
The instrument automatically divides Span into multiple chunks if needed.
1644 Spectrum Analyzer Mode User's &Programmer's Reference