9 Programming the Instrument
9.4 Status Register System & STATus Subsystem
9.4.6.5 Questionable Register
This register and the "Operation Register" on page 2787 monitor the overall
instrument condition. They are accessed using :STATus:OPERation and
:STATus:QUEStionable.
This register monitors the instrument’s condition to see if anything questionable has
happened. It detects anything that might cause an error or a bad measurement,
such as a hardware problem, an out-of-calibration situation, or a unusual signal. All
the bits are summary bits from lower-level event registers.
Bit Condition Operation
3 Power summary Summary bit for "Questionable Power Register" on page 2797
4 Temperature
summary
Summary bit for "Questionable Temperature Register" on page
2800
5 Frequency summary Summary bit for "Questionable Frequency Register" on page
2803
8 Calibration summary Summary bit for "Questionable Calibration Register" on page
2806
9 Integrity summary Summary bit for "Questionable Integrity Register" on page 2819
Filter Registers
"Questionable Condition" on page 2795
"Questionable Enable" on page 2795
"Questionable Event Query" on page 2796
"Questionable Negative Transition" on page 2796
"Questionable Positive Transition" on page 2796