3 Spectrum Analyzer Mode
3.2 Swept SA Measurement
calibration. When the Continue button is pressed, the Open calibration sweep is
taken and stored in internal memory, for use later in this cal process. If the Cancel
button is pressed, the Open/Short Cal is cancelled and the Normalize menu is
returned.
On completion of the Open Calibration, the Short Calibration Form is displayed. This
form shows a diagrammatic representation of how to connect the external source to
the spectrum analyzer to perform the Short calibration. When the Continue button
is pressed, the Short calibration sweep is taken and stored in internal memory, for
use later in this cal process. If the Cancel button is pressed, the Open/Short Cal is
cancelled and the Normalize menu is returned.