3 Spectrum Analyzer Mode
3.4 Occupied BW Measurement
Range
Dual-Attenuator models: Off | Elec Atten Only | Mech + Elec Atten
Single-Attenuator models: Off | On
Backwards Compatibility Command
Notes ON aliases to "Elec Atten Only" (:POW:RANG:OPT:ATT ELEC)
OFF aliases to "Off" (:POW:RANG:OPT:ATT OFF)
:POW:RANG:AUTO? returns true if :POW:RANG:OPT:ATT is not OFF
Backwards
Compatibility
SCPI
[:SENSe]:POWer[:RF]:RANGe:AUTO ON | OFF | 1 | 0
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Adjustment Algorithm
The algorithms for the adjustment are documented below:
742 Spectrum Analyzer Mode User's &Programmer's Reference