3 Using the Oscilloscope
46 Keysight U1610/20A User’s Guide
Adjust the vertical sensitivity
You can increase or decrease the vertical sensitivity (Volt/div) of a waveform by
pressing .
Channel coupling
Toggle to set the channel coupling.
AC coupling blocks any DC component in the waveform and allows only the AC
component of the signal to be viewed.
DC coupling allows both AC and DC components to pass into the handheld scope.
Probe setting
Press to access the probe setup submenu.
Toggle to set the Volts unit for a voltage probe or Amps for a current
probe. Channel sensitivity and offset, trigger level, measurement results, and
math functions will reflect the unit you have selected.
Press repeatedly to set the attenuation factor/sensitivity for
measuring voltage/current values with a voltage/current probe. The attenuation
factor/sensitivity must be set according to the probe being used to ensure the
measurement results reflect the actual voltage/current level.