EasyManua.ls Logo

Keysight U1610A - Capacitance Measurement

Keysight U1610A
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the Digital Multimeter 4
Keysight U1610/20A User’s Guide 95
2 Read the resistance reading from the display.
3 See "Relative Measurement", "Range", and "Restart Measurements" for the
respective functions.
Capacitance Measurement
Capacitance is measured by charging the capacitor with a known current for a
known period of time, measuring the resulting voltage, and then calculating the
capacitance.
To measure capacitance:
1 Press and use the keys to select the capacitance
measurement function. Set up the following connections:
WARNING
Disconnect circuit power and discharge all high voltage capacitors before
performing capacitance measurement to avoid electric shock and damage to
the handheld scope. Use the V DC function to confirm that the capacitor is
fully discharged.