EasyManuals Logo

Keysight U1610A Service Manual

Keysight U1610A
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
8 Keysight U1610/20A Service Guide
Measurement Category
The U1610/20A has a safety rating of CAT II and CAT III.
Measurement CAT I Measurements performed on circuits not directly
connected to the AC mains. Examples are measurements on circuits not derived
from the AC mains and specially protected (internal) mains-derived circuits.
Measurement CAT II Measurements performed on circuits directly connected
to a low-voltage installation. Examples are measurements on household
appliances, portable tools, and similar equipment.
Measurement CAT III Measurements performed in the building installation.
Examples are measurements on distribution boards, circuit- breakers, wiring,
including cables, bus-bars, junction boxes, switches, socket outlets in the fixed
installation, and equipment for industrial use, and some other equipment
including stationary motors with permanent connection to the fixed installation.
Measurement CAT IV Measurements performed at the source of the
low-voltage installation. Examples are electricity meters and measurements on
primary overcurrent protection devices and ripple control units.

Table of Contents

Other manuals for Keysight U1610A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight U1610A and is the answer not in the manual?

Keysight U1610A Specifications

General IconGeneral
BrandKeysight
ModelU1610A
CategoryTest Equipment
LanguageEnglish

Related product manuals