Demodulating AM Signals
Measuring the Modulation Index of an AM Signal
Measuring the Modulation Index of an AM Signal
This procedure demonstrates how to use the signal analyzer as a fixed-tuned
(time-domain) receiver to measure the modulation index as a percent AM value of an
AM signal.
Step Action Notes
1 Set up the signal source. a. Set the source frequency to
300 MHz.
b. Set the source amplitudes to
10 dBm.
c. Set the AM depth to 80%.
d. Set the AM rate to 1 kHz.
e. Turn AM on.
2 Connect an Keysight
ESG RF signal source to
the analyzer RF INPUT
as shown.
3 Select the mode. Press Mode, Spectrum
Analyzer.
4 Preset the analyzer. Press Mode Preset.
5 Set the center frequency,
span, RBW and the
sweep time.
a. Press FREQ Channel,
Center Freq, 300, MHz.
b. Press SPAN X Scale, Span,
500, kHz.
c. Press BW, Res BW, 30,
kHz.
d. Press Sweep/Control,
Sweep Time, 20, ms.

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals