EasyManuals Logo

Keysight x-series Measurement Guide

Keysight x-series
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #220 background imageLoading...
Page #220 background image
220
Concepts
Occupied Bandwidth Measurement Concepts
Occupied Bandwidth Measurement Concepts
Purpose
Occupied bandwidth measures the bandwidth containing 99.0 of the total
transmission power.
The spectrum shape of a signal can give useful qualitative insight into transmitter
operation. Any distortion to the spectrum shape can indicate problems in transmitter
performance.
Measurement method
The instrument uses digital signal processing (DSP) to sample the input signal and
convert it to the frequency domain. With the instrument tuned to a fixed center
frequency, samples are digitized at a high rate with DSP hardware, and then
converted to the frequency domain with FFT software.
The total absolute power within the measurement frequency span is integrated for its
100% of power. The lower and upper frequencies containing 0.5% each of the total
power are then calculated to get 99.0% bandwidth.

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals