Making Time-Gated Measurements
Gated LO Measurement
Gated LO Measurement
This procedure utilizes gated LO to gate the FM signal. For concept and theory
information about gated LO see “How time gating works” on page 199.
Step Action Notes
1 Set the analyzer to the
Spectrum Analyzer
mode.
Press Mode, Spectrum
Analyzer.
This enables the spectrum analyzer
measurements.
2 Preset the analyzer. Press Mode Preset.
3 Set the center frequency,
span, reference level and
attenuation.
a. Press FREQ Channel,
Center Freq, 40, MHz.
b. Press SPAN X Scale,
Span, 500, kHz.
c. Press AMPTD Y Scale,
Ref Level, 10, dBm.
4 Set the gate source to the
rear external trigger
input.
.Press Sweep/Control,
Gate, More, Gate
Source, External 1.
5 Set the gate delay, gate
length, gate sweep time,
and gate trigger.
a. Press Sweep/Control,
Gate, Gate Delay, 2,
ms.
b. Press Gate Length, 1,
ms.
c. Press Gate View Sweep
Time, 5, ms.
d. Press More1 of 2,
Control (Edge).
6 Access the analyzer gate
view display.
Press Sweep/Control,
Gate, Gate View (On).
Use this function to confirm the gate “on”
time during the RF burst interval
(alternatively you could also use the
oscilloscope to view the gate settings).

Table of Contents

Other manuals for Keysight x-series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight x-series and is the answer not in the manual?

Keysight x-series Specifications

General IconGeneral
Analysis BandwidthUp to 1 GHz (depending on model)
Real-Time BandwidthUp to 1 GHz (depending on model)
Form FactorBenchtop
ConnectivityLAN, USB, GPIB
Resolution Bandwidth1 Hz to 10 MHz (depending on model)

Related product manuals