88 88 V V 1.01.0
Service ManualService Manual
Operating Lights marLEDOperating Lights marLED
®®
E9 / E9i / E15E9 / E9i / E15
b) Functional check according to Instructions for Useb) Functional check according to Instructions for Use
Perform a functional check of all operating functions of the light (combination). For thisPerform a functional check of all operating functions of the light (combination). For this
purpose, also observe the Instructions for Use for the respective light type and thepurpose, also observe the Instructions for Use for the respective light type and the
Instructions for Use for auxiliary equipment (e. g. monitors, camera).Instructions for Use for auxiliary equipment (e. g. monitors, camera).
√√
c) Checking the light fieldc) Checking the light field
Align the lamp body horizontally and check if the lamp produces a round light field whenAlign the lamp body horizontally and check if the lamp produces a round light field when
holding it one meter away from the ground.holding it one meter away from the ground.
√√
d) Checking the light intensityd) Checking the light intensity
Measure the light intensity with the smallest light field and without dimming at a distanceMeasure the light intensity with the smallest light field and without dimming at a distance
of one meter to the ground.of one meter to the ground.
Nominal value ranges of the light intensityNominal value ranges of the light intensity
Model marLED
Model marLED
®®
E9 E9 marLEDmarLED
®®
E15E15
E E c c 100–120 100–120 klx klx 135 135 to to 160 160 klxklx
Note down the determined value in the inspection protocol.Note down the determined value in the inspection protocol.
√√
e) Checking the lamp voltage at nominal load on e) Checking the lamp voltage at nominal load on the contact blockthe contact block
The supply voltage that is measured on the contact block (at maximum brightness setting)The supply voltage that is measured on the contact block (at maximum brightness setting)
must be at least 26 VDC and must not exceed 36 VDC.must be at least 26 VDC and must not exceed 36 VDC.
√√