Service ManualService Manual
Operating Lights marLEDOperating Lights marLED
®®
E9 / E9i / E9 / E9i / E15E15
4.) Inspection of the lamp body4.) Inspection of the lamp body
a) Checking for damagea) Checking for damage
•• Check the lamp body for damage / cracks.Check the lamp body for damage / cracks.
••
√√
b) Checking the connections and the b) Checking the connections and the plug contactsplug contacts
•• Check all plug connections and cable connections on the boards (see 6.13.1 “ConnectionsCheck all plug connections and cable connections on the boards (see 6.13.1 “Connections
for Supply Voltage and Light Engine”, page 49 for the control electronics. They may havefor Supply Voltage and Light Engine”, page 49 for the control electronics. They may have
to be fixed using cable ties.to be fixed using cable ties.
••
√√
c) Installing current softwarec) Installing current software
••
Observe section 6.19 “”, page 55. The new software version must now be noted on theObserve section 6.19 “”, page 55. The new software version must now be noted on the
protocol and the control board.protocol and the control board.
••
√√
d) Checking the Light Engines and ground glassd) Checking the Light Engines and ground glass
•• Check the Light Engines and the inside of the ground glass for soiling and damage.Check the Light Engines and the inside of the ground glass for soiling and damage.
Aggressive cleaning agents must not be used for cleaning purposes.Aggressive cleaning agents must not be used for cleaning purposes.
••
√√
e) Checking the lamp casing and back e) Checking the lamp casing and back plateplate
••
Check the lamp casing and the rear plate for soiling, damage and damage to paintwork.Check the lamp casing and the rear plate for soiling, damage and damage to paintwork.
Aggressive cleaning agents must not be used for cleaning purposes.Aggressive cleaning agents must not be used for cleaning purposes.
••
√√
5.) Inspection of the light functions5.) Inspection of the light functions
Tools:Tools:
Instructions for Use, voltmeter True-RMS voltmeter, luxmeter,Instructions for Use, voltmeter True-RMS voltmeter, luxmeter,
measuring range 0.1 to 199 900 lxmeasuring range 0.1 to 199 900 lx
a) Checking the emergency power switch of KLS Martin from mains to a) Checking the emergency power switch of KLS Martin from mains to batterybattery
••
Check the correct function of the emergency power switch (if applicable), for example byCheck the correct function of the emergency power switch (if applicable), for example by
tripping the circuit breaker of the respective light (combination) in the building or bytripping the circuit breaker of the respective light (combination) in the building or by
removing the fuse holder on the series terminal of the ceiling tube.removing the fuse holder on the series terminal of the ceiling tube.
••
√√