Service ManualService Manual
Operating Lights marLEDOperating Lights marLED
®®
E9 / E9i / E15E9 / E9i / E15
Guidance and manufacturer’s declaration according to
Guidance and manufacturer’s declaration according to IEC 60601-1-2,IEC 60601-1-2,
clause 6.8.3.201 b) Table 204: Electromagnetic immunityclause 6.8.3.201 b) Table 204: Electromagnetic immunity
The device is intended for The device is intended for use in an use in an electromagneelectromagnetic environment as specified below. The user tic environment as specified below. The user shouldshould
assure that it is used in such an assure that it is used in such an environmenenvironment.t.
Immunity testImmunity test IEC 60601IEC 60601
test leveltest level
ComplianceCompliance
levellevel
Electromagnetic environment–GuidanceElectromagnetic environment–Guidance
Conducted RFConducted RF
according toaccording to
IEC 61000-4-6IEC 61000-4-6
Radiated RFRadiated RF
according toaccording to
IEC 61000-4-3IEC 61000-4-3
3 V3 V
effeff
150 kHz to150 kHz to
80 MHz80 MHz
3 V/m3 V/m
80 MHz to80 MHz to
2.5 GHz2.5 GHz
10 V10 V
effeff
150 kHz to150 kHz to
80 MHz80 MHz
10 V/m10 V/m
80 MHz to80 MHz to
2,5 GHz2,5 GHz
Portable and mobile Portable and mobile RF communicationsRF communications
equipment should be used no closer to any part ofequipment should be used no closer to any part of
the device (including cables) than thethe device (including cables) than the
recommended separation distance calculated fromrecommended separation distance calculated from
the equation applicable to the frequency of the equation applicable to the frequency of thethe
transmitter.transmitter.
RecommenRecommended ded separation distance:separation distance:
PPdd 3535..00==
PPdd 3535..00==
for 80 MHz to 800 MHzfor 80 MHz to 800 MHz
PPdd
7070..00==
for 800 MHz to 2,5 for 800 MHz to 2,5 GHzGHz
wherewhere PPis the maximum output power rating ofis the maximum output power rating of
the transmitter in watts (W) according to thethe transmitter in watts (W) according to the
transmitter manufacturer andtransmitter manufacturer and ddis theis the
recommended separation distance in meters (m).recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, asField strengths from fixed RF transmitters, as
determined by an electromagnetic site surveydetermined by an electromagnetic site survey
aa
,,
should be less than the should be less than the compliance level in eachcompliance level in each
frequency range.frequency range.
bb
Interference may occur in the vicinity ofInterference may occur in the vicinity of
equipment marked with the equipment marked with the following symbol.following symbol.
Note 1:Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:Note 2:
These guidelines may not apply in These guidelines may not apply in all situations. Electromagnetic propagatioall situations. Electromagnetic propagation isn is
affected by absorption and reflection from structures, objects and affected by absorption and reflection from structures, objects and people.people.
aa
Field strengths from fixed transmitters, such as base Field strengths from fixed transmitters, such as base stations for radio (cellular / stations for radio (cellular / cordless) telephonescordless) telephones
and land mobile radios, amateur radio stations and AM and FM radio and TV and land mobile radios, amateur radio stations and AM and FM radio and TV stations, cannot bestations, cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RFRF
transmitters, an electromagnetic site survey should be considered. If the transmitters, an electromagnetic site survey should be considered. If the measured field strength inmeasured field strength in
the location in which the the location in which the device is used exceeds the applicable RF compliance level specified above, thedevice is used exceeds the applicable RF compliance level specified above, the
device should be observed to verify normal device should be observed to verify normal operation. If abnormal performance is observed, additionaloperation. If abnormal performance is observed, additional
measures may be necessary, such as measures may be necessary, such as reorienting or relocating the device.reorienting or relocating the device.
bb
Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m.Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m.