9. The first screen allows you to set the Type of cabinet. The only difference
between the Type A1 and Type A2 configuration setting is that the airflow
velocity limits for normal operation of the optional airflow
sensor (because the Type A1 has a nominal inflow of 85
FPM, and the Type A2 is 105 FPM). If the air velocity goes
out of range, the alarm will be displayed, but the cabinet will
continue to run. Selecting Type B2 will cause an inflow
alarm if the inflow velocity drops below the alarm setpoint,
and will result in the cabinet blower stopping. Use the
or buttons to select the cabinet Type, and then press
OK/Mute.
10. The second screen allows you to configure the cabinet for a
canopy or not. Select the “a canopy” option.
11. This screen configures the unit for a UV light or not. After
your selection is chosen, and you press OK/Mute, the unit
will shut off and then reset in the new configuration.
Configuration
Type A Canopy Option
This cabinet has:
or to choose,
OK to select, MENU to escape
a canopy
no canopy


Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30132 and is the answer not in the manual?

Labconco Purifier Logic+ 30132 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30132
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals