Product Service 1-800-522-7658
145
Calibration Procedure-
NOTE: YOU MUST FOLLOW THESE INSTRUCTIONS EXACTLY, OR THE SENSOR
WILL NOT OPERATE PROPERLY!
Note: You will need to access the certifiers menu to calibrate the airflow
sensor.
Note: To properly calibrate the air flow sensor, a qualified certifier must
verify the inflow and downflow values.
With the cabinet operating:
8. Press the
Menu
button.
9. Press the button until the
Services
option is highlighted
(it will turn yellow).
10. Press
OK/Mute
to enter the Services menu screen.
11. Press the button until the
Calibration
option is
highlighted (it will turn white).
12. Press
OK/Mute
to start the calibration procedure.
13. You will see a
warning screen
alerting you that you are
about to alter the BSC’s settings.
14. Press
OK/Mute
to input the password.
NOTE: IF THE UNIT IS CONNECTED TO AN EXHAUST SYSTEM, YOU MUST
STOP ALL AIR FLOW OVER THE SENSOR DURING THE INITIAL ZEROING
OF THE SENSOR - DO THIS BY SHUTTING OFF THE EXHAUST SYSTEM, OR
BLOCKING THE SENSOR INTAKE WITH TAPE.
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Configuration
Calibration

Service
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Calibration
OK to proceed, MENU to return
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape