3.8.3 Defining Kinetic Measurements
The following example describes a kinetic measurement of a 96-well microplate.
1. Select a 96-well microplate from the Plate definition drop-down list.
2. Insert a Kinetic Cycle program element between the part of plate and
the absorbance element.
3. Cycles/Number of cycles: 50
4. Kinetic interval (interval between measurements): select Use kinetic
interval and enter: 2 minutes 30 seconds.
5. Define the Absorbance element by entering the desired measurement
wavelength.