EasyManua.ls Logo

Labtech LT-4500 - Instrument Linearity with Liquid Samples

Labtech LT-4500
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5. Quality Control
47
5.3.4 Instrument Linearity with Liquid Samples
The linearity for the instrument and application at the wavelength used can be
checked by using a dilution series of a solution.
The result depends on the purity of the dye used and the meniscus of the liquid in
the wells.
As a reference, a dilution series of Orange G solution for measurements at
492 nm can be used.
The dilution series should be within the range of 0.1 to 3.0 OD. To reach about
3 OD it is recommended to use 125 mg.l
-1
Orange G (Sigma, Cat. No. O7252).
For other wavelengths, different solutions must be used.
200 µl of each dilution are then pipetted into the microplate, a minimum of at least
two samples should be used for each dilution to reduce the errors caused by
pipetting.
The microplate is then measured and a linear regression of OD against
concentration is drawn from the average of the measured OD values.
Determine the unweighted residual square value R
2
of the regression line.
Typical residual square values for a standard application are equal or better than
R
2
= 0.998.
Note
Data can vary due to pipetting inaccuracy.

Table of Contents