EasyManua.ls Logo

Labtech LT-4500 - Measurement Specifications

Labtech LT-4500
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Instrument Features
40
4.4.2 Measurement Specifications
PARAMETERS CHARACTERISTICS
Measurement time:
single wavelength
dual wavelength
< 15 seconds
< 20 seconds
Wavelength range:
Standard
400 - 750 nm
Measurement range:
400 - 750 nm
0 - 4.000 OD
Resolution: 0.0001 OD
Accuracy: 450, 492 nm
0.000 - 2.000 OD
2.000 - 3.000 OD
(1.0 % + 0.010 OD)*
(1.5 % + 0.010 OD)*
Precision: 450, 492 nm
0.000 - 2.000 OD
2.000 - 3.000 OD
(0.5 % + 0.005 OD)*
(1.0 % + 0.005 OD)*
Linearity: 450, 492 nm
0.000 - 2.000 OD
2.000 - 3.000 OD
1 %
1.5 %
Wavelength selection:
Standard filter
Narrow bandwidth interference
filters. Up to 8 filters can be
mounted in a filter wheel.
Filter wavelength accuracy: Central wavelength ± 2 nm
Filter bandwidth:
At 50 % transmission
10 ± 2 nm
Light source: LED
Computer interface:
USB port
USB 1.1 / 2.0
All connected devices must be approved and listed as per IEC 60950-1
Information Technology Equipment Safety and equivalent local standards.
* better than or equal to x % of measurement value plus corresponding OD value.

Table of Contents