4. Instrument Features
35
4. Instrument Features
4.1 Instrument Features
The following absorbance measurement modes are available on the LT-4500:
endpoint, kinetic, and multilabel measurements.
4.1.1 Microplate Shaking
The LT-4500 is able to shake the microplate before it is measured. The
microplate can also be shaken between each of the kinetic measurement cycles.
Use LT-com to set the shaking mode.
CAUTION
WHEN SHAKING MICROPLATES, SPILLAGE MAY OCCUR IF THE
WELLS ARE FILLED WITH A TOO HIGH VOLUME.
Shaking modes for the LT-4500:
Shake Mode Shake Width Shake Frequency
HIGH
NORMAL
LOW
WIDE
2.8 mm
4.4 mm
4.4 mm
14.1 mm
12.3 Hz
9.2 Hz
7.8 Hz
2.0 Hz