7. Cleaning, Maintenance, and Disposal
72
5. Carefully apply the disinfectant solution on the plate transport according
to the manufacturer’s Instructions for Use.
Do not use too much disinfectant to prevent the solution flowing into the
instrument or soiling the lenses when moving the plate transport into the
device.
6. After the required contact time (according to the manufacturer’s
Instructions for Use) wipe the plate transport using a soft paper towel
moistened with a mild detergent or distilled water to remove all traces of
the disinfectant.
7. Carefully move the plate transport into the instrument.
8. Carefully apply the disinfectant on the base plate of the plate transport.
9. After the required contact time wipe the base plate of the plate transport
using a soft paper towel moistened with a mild detergent or distilled water
to remove all traces of the disinfectant