2/19/98 Log - Logging Sound Level Meter (Option 01) 6-49
Step 24 For this example, we will use “Level”. Press
u or d to highlight “Level”. Then, press
c to activate this selection.
Step 25 Now, press u or d to highlight “Excd
Time”.
Step 26 Then, press c to enter the “Excd Time”
menu.
Here you establish the criteria for
where the exceedance time occurred.
“Start” logs the exceedance time as
the moment when the sound level
crossed the predefined exceedance
level. “Max” logs the exceedance
time at the point where the maxi-
mum sound level occurs during the
exceedance event.
Step 27 Use u or d to highlight “Start”. Then,
press c to activate this selection.