X-Stream Operator’s Manual
WM-OM-E Rev I 253
.
Touch Screen Calibration
Touching the Touch-Screen Calibration button starts the calibration procedure. During
the procedure, you will be prompted to touch the center of a small cross in 5 key locations on the
touch screen. Because sufficient accuracy cannot be achieved using your finger, use a stylus
instead for this procedure. The calibration has a ten-second timeout in case no cross is touched.
To avoid parallax errors, be sure to place your line of sight directly in front of each cross before
touching it.