396 WM-OM-E Rev I
ltpt Local Time Peak-to-Trough
Definition
The time between a local feature peak and trough.
Description
The average of the time between all local feature peaks and troughs is
displayed as ltpt. Provided for histograms are the times between
peak-trough pairs for all local features between the parameter cursors.
Parameter Settings
Selecting this parameter from the Measure menus prompts you for a
hysteresis setting, which allows you to set the hysteresis value to a
specified number of vertical divisions.
Example