Safety Notes
30
Leica HM500/Ref. 10 715 697/Version B
Responsibility
As a superior legal entity, the owner is responsible for the
product being used and maintained as intended, for providing
information to other users, and for assigning authority to
o
perate the product. Furthermore, the use of other devices or
items of equipment in combination with the Leica HM500 also
falls within the responsibility of the owner.
The Leica HM500 may be operated by trained and qualified
persons only. Persons working with this device or with
accessories of this device must not fail to read the user man-
ual and the owner must ensure that users have understood
the contents of these instructions.
Leica Microsystems (Schweiz) AG cannot accept any respon-
sibility or liability whatsoever for any direct or indirect dam-
age or loss arising from the use or application of the
Leica HM500.
Legal regulations
National and local generally applicable safety and accident
prevention laws and regulations must be complied with.
Leica Microsystems (Schweiz) AG points out that the
Leica HM500 like any other operated medical product is
subject to regular mandatory safety checks by the owner
in accordance with the relevant national medical product
legislation or other national laws or ordinances. Should you
have any questions in this regard, contact your dealer or
Leica Microsystems (Schweiz) AG.
Information for the user
Follow the instructions given in this manual.
Follow the instructions given by your employer regarding the
organization of work and safety at work.
The user manual must be kept in the immediate vicinity of
the device to ensure that it can be referred to by the user if
need be.
In the event of malfunctions (a blurred image or none at all)
stop using the Leica HM500.
If the device is dropped or has any visible mechanical dam-
age (for example, to the housing or to the height adjustment)
or any other damage, the device should not be used under
any circumstances!
In the event of problems or if in doubt, please contact your
dealer!
The specified IP X1 protection of the footswitch in accordance
with IEC 60529 applies when the footswitch is plugged in.
Notes on the Leica HM500 trolley
Make sure that all equipment placed on the Leica HM500
trolley is securely stored.
It is advisable to use two people to place all heavy compo-
nents on the Leica HM500 trolley. This is particularly recom-
mended in the case of the monitor.
Electrical equipment should not be operated in the vicinity of
gases such as flammable gas used in anesthesia or similar
gases.
Do not exceed the load rating of the Leica HM500 trolley.
Observe the maximum load capacity of the Leica HM500
trolley (see Technical Data).
The Leica HM500 trolley is intended to provide a practical and
mobile workstation for medical electrical equipment. For the
entire medical electrical system to be compliant with
IEC 60601-1-1, the total ground leakage current must not
exceed the maximum limit of 0.5 mA. If the total ground leak-
age currents exceed this tolerance threshold, the system will
have to be powered via a safety isolating transformer. If an
isolating transformer is not installed, the multiple socket out-
let of the Leica HM500 trolley must not be used for connect-
ing equipment which is not in conformity with the require-
ments for ground leakage currents under IEC 60601-1-1. If an
isolating transformer is installed, the total power uptake of all
connected units must remain within the rated value for the
transformer.
If potential equalization is necessary, first connect the
optional POAG connector of the trolley to the room’s POAG
connector. Next, connect the POAG plug on the Leica HM500
trolley with the POAG pin of the instrument.