GA05141_0702 - 09/2004
102
Troubleshooting
Measures
Check pump (temperature sensor). For measures
see failure Bearing Temp.
Check cable conections.
Check if the Pt 100 is connected to the corre-
sponding sensor cable. The sensor cables are
marked with COOLING WATER or BASFLANGE
(TMS-sensor).
Check cooling temperature sensor.
For Pin assignment see Fig. 44.
Measure the resistance between pins 1 and 3 of
the cooling temp. sensor.The resistance is typical-
ly between 110 and 150 Ohm (20°C to 140°C).
Contact Leybold Service if the sensor is defective.
Check cable connections.
Check cable. Replace the cable if it is damaged
Contact Leybold service. Principally controllers
with SR-Software Revision 302.18 or higher can
be updated for new pumps (see 3.10.2 for Conv.
SR SW-Rev).
Contact Leybold service.
Contact Leybold service.
Contact Leybold service.
Contact Leybold service.
Contact Leybold service.
Check cable connections.
Check cable. Replace cable if it is damaged.
Contact Leybold service.
Failure Message on
Display
Bearing Temp. open
Sensor loop is interrupted
Cooling Temp. open
Sensor loop is interrupted
AMB Not Initial
Converter can not identify the
pump.
SPI Com.-Fail
Communication problem bet-
ween main controller and magne-
tic bearing controller
SPI-Timeout
Communication problem bet-
ween main controller and magne-
tic bearing controller
Bearing Overload
Magnetic Bearing output current
is overloaded
Internal Overload
DC/DC Converter is overheated
Rotor Not Lifted
PK-Communication
Converter does not communicate
with the memory chip of the
pump.
The failure occurs when the fre-
quency is lower than 5 Hz.
In case of a frequency above 5
Hz a warning will occur.
Possible Cause
Pt 100 (temperature sensor bea-
ring) damaged.
Pt 100 (temperature sensor coo-
ling) is not connected.
Pt 100 damaged.
Drive / Bearing Cable is not con-
nected with the pump, pumps
memory chip "PK-..” or the con-
troller.
Drive / Bearing Cable or connector
damaged
Controller is not able to operate
with the connected pump
Converter failure
Converter failure
Converter failure
Converter failure
Converter failure
Drive / Bearing Cable is not con-
nected with the pump, pumps
memory chip "PK-..” or the con-
troller.
Drive / Bearing Cable or connector
damaged.
PK. or converter failure

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LEYBOLD MAG Series and is the answer not in the manual?

LEYBOLD MAG Series Specifications

General IconGeneral
ApplicationsLaboratory, medical, and industrial applications
DriveMagnetic drive
SealingHermetically sealed
Flow RateVaries depending on the model, ranging from a few liters per minute (LPM) to several tens of LPM. Consult specific model datasheets for precise values.
Max PressureVaries by model (typically up to 10 bar)
Power SupplyTypically available in various AC voltages (e.g., 115V, 230V) and frequencies (50/60 Hz)
Power ConsumptionVaries depending on the model, ranging from a few watts to several hundred watts. Consult specific model datasheets for precise values.
Operating TemperatureTypically 0°C to +60°C
WeightVaries depending on the model, ranging from a few kilograms to several kilograms. Consult specific model datasheets for precise values.
DimensionsVaries depending on the model. Consult specific model datasheets for precise dimensions.
MaterialsStainless steel