Operation
3.2 Temperature Management
System
Function description
The TMS is only in function, when a MAG CT version is connected. The
heater will be activated when the mains is switched on. The TMS controls
the heating and water cooling to maintain the pump at the specified setpo-
int temperature.
In order to maintain an almost homogeneous temperature distribution in the
pump the system is equipped with several temperature sensors. The mea-
sured values of these sensors are used to determine two temperatures for
the TMS: T
TMS
and T
A
. Both temperatures are used to switch on and off the
heater and the cooling water valve.
Switching points
Heater OFF ON T
TMS
< T
set
– 1 K
Heater ON OFF T
TMS
> T
set
+ 1 K
Cooling OFF ON T
A
> T
set
+ 2 K
Cooling ON OFF T
A
< T
set
+ 1 K
TMS ok T
set
– 5 K T
TMS
T
set
+ 5 K
Warning TMS T
TMS
> T
set
+ 5 K
Settings
The temperature (TMS) setpoint can be programmed via the front panel
keys at every time.
Principally the factory presetting will be used. The setting is saved in the
pump’s memory chip. Before changing any setpoint value request Leybold!
For the setting refer to 3.10.6 Operating menu, Set TMS.
Heat up
The setpoint temperature will be reached within 30 to 60 minutes depen-
ding on cooling and environmental conditions. For temperature sensitive
applications observe the cooling water specifications (refer to Section 1.6
and 2.5).
In order to guarantee correct temperature setting of the pump it is
required to provide the cooling water within the envelope descri-
bed in Section 2.5.
Caution
only for MAG ... CT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LEYBOLD MAG Series and is the answer not in the manual?

LEYBOLD MAG Series Specifications

General IconGeneral
ApplicationsLaboratory, medical, and industrial applications
DriveMagnetic drive
SealingHermetically sealed
Flow RateVaries depending on the model, ranging from a few liters per minute (LPM) to several tens of LPM. Consult specific model datasheets for precise values.
Max PressureVaries by model (typically up to 10 bar)
Power SupplyTypically available in various AC voltages (e.g., 115V, 230V) and frequencies (50/60 Hz)
Power ConsumptionVaries depending on the model, ranging from a few watts to several hundred watts. Consult specific model datasheets for precise values.
Operating TemperatureTypically 0°C to +60°C
WeightVaries depending on the model, ranging from a few kilograms to several kilograms. Consult specific model datasheets for precise values.
DimensionsVaries depending on the model. Consult specific model datasheets for precise dimensions.
MaterialsStainless steel