Lighthouse HANDHELD 2016, 3016, 5016 Gen F Operating Manual
C-4
248083400-1 Rev 7
Registers 40200-40999 are reserved for eight character names
40034 unsigned integer Sample Time [low]
40035 unsigned integer Data Set [high]. Works in conjunction with 40036. Data entered
here is applied to the device through the command register.
40036 unsigned integer Data Set [low]
40037 unsigned integer Alarm Mode. Type of alarming performed
40038 unsigned integer Alarm Parameter. Control parameter for given alarm mode.
40039 unsigned integer Laser Reference Voltage (millivolts)
40040 unsigned integer View Volume. Divide by 100 to get percentage.
For example: 6550d = 65.50%
40041 ASCII string Flow Unit. Defines unit as cfm, lpm, mlpm.
40042 ASCII string Flow Unit. char[2], char[3]
40043 unsigned integer Calibration Reference Voltage (millivolts)
40047 signed integer Calibration Due Date [high]. Indicates when instrument is due
for calibration. this number can be generated by the ANSI C/
C++ time() function.
40048 signed integer Calibration Due Date [low].
...
40199 unsigned integer Number of available alphanumeric location names (0 means
alphanumeric names are not supported).
40200 ASCII string Location_1_char[0], char[1] (NULL terminated string)
40201 ASCII string Location_1_char[2], char[3]
40202 ASCII string Location_1_char[4], char[5]
40203 ASCII string Location_1_char[6], char[7]
...
40996 ASCII string Location_200_char[0], char[1] (NULL terminated string)
40997 ASCII string Location_200_char[2], char[3]
40998 ASCII string Location_200_char[4], char[5]
40999 ASCII string Location_200_char[6], char[7]
Table C-3 Sensor Settings Registers
Register Data Type Description