HANDHELD MODBUS Register Map v1.48
248083400-1 Rev 7
C-11
Setting the Alarm Threshold Value
The Alarm Threshold Value is set in the low register of the channels.
45014 unsigned int Threshold for Particle Channel 3 [low]
45015 unsigned int Threshold for Particle Channel 4 [high]
45016 unsigned int Threshold for Particle Channel 4 [low]
45017 unsigned int Threshold for Particle Channel 5 [high]
45018 unsigned int Threshold for Particle Channel 5 [low]
45019 unsigned int Threshold for Particle Channel 6 [high]
45020 unsigned int Threshold for Particle Channel 6 [low]
Table C-10 Alarm Threshold Register Default Values
Registers
Particle
Channel
Threshold
Value
45009 - 45010 1 1000
45011 - 45012 2 1000
45013 - 45014 3 1000
45015 - 45016 4 1000
45017 - 45018 5 1000
45019 - 45020 6 1000
Table C-9 Alarm Threshold Registers