HANDHELD MODBUS Register Map v1.48
248083400-1 Rev 7
C-9
Enable Alarming for a Channel
To enable alarming on the third particle channel, the user would enable
Bit 1 for register 43014.
To disable alarming on the third channel and enable alarming on the
second channel, disable Bit 1 for register 43014 and enable Bit 1 for
register 43012.
To disable alarming completely, disable Bit 1 for register 43012. Now,
no channels are enabled for alarms.
43016 unsigned int Enable for Particle Channel 4 [low]
43017 unsigned int Enable for Particle Channel 5 [high]
43018 unsigned int Enable for Particle Channel 5 [low]
43019 unsigned int Enable for Particle Channel 6 [high]
43020 unsigned int Enable for Particle Channel 6 [low]
...
43041 unsigned int Enable for Analog Channel 1 [high]
43042 unsigned int Enable for Analog Channel 1 [low]
43043 unsigned int Enable for Analog Channel 2 [high]
43044 unsigned int Enable for Analog Channel 2 [low]
43045 unsigned int Enable for Analog Channel 3 [high]
43046 unsigned int Enable for Analog Channel 3 [low]
43047 unsigned int Enable for Analog Channel 4 [high]
43048 unsigned int Enable for Analog Channel 4 [low]
Table C-7 Alarm Enable Registers
Register Data Type Description
Table C-8 Example of Alarming on Channel 2
Registers
Particle
Channel
Bit 1
Enabled
43009 - 43010 1 0
43011 - 43012 2 1