Operation
248083400-1 Rev 7
4-63
• Press MAIN to return to the MAIN screen.
• Press CFG to enter the Configuration screen.
• Press SAMPLE to set the Sample time.
• Press the VOLUME button to set the instrument’s sample volume.
• Type in 146 and press ENTER. This sets the instrument for the
channel size with the largest minimum volume requirement.
Note:
The Sample time
automatically changes to 1:28
(1 minute, 28 seconds).
• Press the CYCLES button.
• Set the CYCLES value for 1 and press ENTER.
Note:
This is all that is
needed to meet the minimum
requirements to classify the
report.
Increase any of the
configuration parameters in
order to meet additional
reporting needs.
• Press the MAIN button.
• Press the Location Select button at the top of the screen.
• Select the location where the first measurement will be taken.
• Press START. Let the instrument record 1 sample at the first
location.
• After the instrument has recorded the sample, change the Location
using the Location Select button.
• Then move the instrument to the second location and press START.
• After the instrument has recorded the second sample, continue to
change the Location setting before recording at the next location,
then move it to the third and forth locations, taking at least one
sample at each location. At one location take 2 samples.
• After the instrument finishes recording at the last location, and the
instrument has a total of at least 5 samples, the instrument is ready
to print the report.
Report Output Setup
Note: Report Output can
also be configured as a
new or previously saved
recipe, which can be set
through the Reports option
or the Recipe option.
The Report output can be directed to an external thermal printer. To
configure the destination, press CFG then OUTPUT. In the Output
Setup screen, verify that the Output on Sample is enabled with a green
check mark.
Printing a Report
To print a report, the user can choose to either clear the data before
starting the test or run the samples and then select the range of data to
include in the report.
• Once the user has collected the necessary samples, press REPORTS
from the MAIN screen.
• Press the RANGE button on the REPORTS screen
• Using the ARROW buttons, scroll to the first record in the range of
data for the report.
• Press the RANGE button on the Data View Screen.
• Type in the number of records to be included in the range.
• Press ENTER.