4 Operation
This chapter describes how to use the HANDHELD 2016, 3016 and
5016 Airborne Particle Counters.
Using the
Instrument
The HANDHELD comes with a charged battery and is ready for use.
To start using the instrument, proceed as follows:
WARNING:
Do not
attempt to sample
reactive gases (such as
hydrogen or oxygen)
with this instrument.
Reactive gases create an
explosion hazard in the
instrument.
Sampling any gas under
pressure can damage the
instrument and void the
warranty.
Sampling any gas that is
not the same density as
ambient air can result in
inaccurate data.
Contact Lighthouse for
more information.
1. Insert the battery included with the shipment. (See Chapter 2 for
instructions.)
2. Position the instrument in the environment to be measured.
3. Remove the protective cap from the inlet tube. To use the
provided isokinetic probe, install it by connecting to inlet tube on
the top of the instrument. NOTE: Do not discard the protective
cap. It should be placed on the inlet tube any time the
instrument is moved outside the environment being
measured.
4. The included Temperature/Relative Humidity probe can be
attached to the provided receptacle to read environmental data.
5. Set on/off switch found on the left side of the unit to ON.
6. The Start Up screen displays on the LCD.
WARNING:
Do NOT
allow water, solvents, or
other liquids enter the
instrument via the inlet
tube - the instrument will
be damaged and void the
warranty.
Do NOT operate the
instrument with the inlet
tube capped or plugged -
the internal pump will be
damaged and void the
warranty.
7. The MAIN screen appears.
8. On the touch screen, press the START button to start the
instrument.
9. “STARTING” will display when the pump is initially turned on.