Specifications
Size Range 0.2 - 10.0 μm
HANDHELD 2016
Channel Thresholds
0.2, 0.3, 0.5, 0.7, 1.0, 2.0 μm
HANDHELD 3016
Channel Thresholds
0.3, 0.5, 0.7, 1.0, 2.5, 5.0 μm
HANDHELD 5016
Channel Thresholds
0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm
Other sizing available for each model; specify
at time of order
Flow Rate 0.1 CFM (2.83 LPM)
Counting Efficiency 50% (per ISO 21501-4)
Laser Source Laser diode
Zero Count Level <1 count/5 minutes (per ISO 21501-4)
Calibration NIST Traceable
Count Modes Concentration, manual/automatic, beep,
cumulative/differential
Data Storage Up to 3000 sample records, includes particle
& environmental data, plus location and time
Communication Modes RS232 or RS485 via RJ-45 to PC or printer
Supporting Software LMS XChange Data Transfer Software, LMS
Express/RT/RT Plus; LMSNet
Environmental Sensors Temperature/Relative Humidity Probe:
0-150°F (-17.8 to 65.6°C) ±1.8°F @ 77°F, 0-
100% ± 5% @ 33%
Touch Screen Display 3.8" (9.25 cm), 320x240
Printer External thermal printer (optional)
Reports FS-209E (ft) , FS-209E (m), ISO 14644-1, EU
GMP (0.5 and 5.0μ channels must be active)
Key Software Features Historical data review, password protection
Enclosure High impact injection molded plastic
Sample Output Internally filtered to HEPA standards
(>99.97% @ 0.3μm)
Vacuum Source Internal pump, flow controlled
Table 2-1 Specifications