HANDHELD MODBUS Register Map v1.48
248083400-1 Rev 7
C-7
11 Instrument Start (Automatic Counting).
Particle Counters: Uses defined Initial Delay, Hold Time, Sample
Interval and counting mode. Instrument executes samples and holds
until an Instrument Stop command is issued. For instruments with
pumps, this command will start the pump.
Manifold Controller
: Uses defined Manifold Sequence. Stops
counting and changing positions when an Instrument Stop
command is issued.
12 Instrument Stop. Aborts current sample. Stops pump, if applicable.
Stops data collection.
13 Set Real Time Clock.
Writes “Data Set” values (from Registers 40035 & 40036) to the
local Real Time Clock. New time value is saved.
192 Change BAUD to 19200
576 Change BAUD to 57600
1152 Change BAUD to 115200
Table C-5 Command Register
Value Action