The MAIN screen displays the following options and information.
• LOCATION: Displays the location that is currently being
measured. Up to 200 alphanumeric locations can be configured.
• LOCATION SELECT button: Allows user to change location
before sampling.
• RECIPE button: Allows the user to view, load and unload recipes
that have been configured and saved in the recipe data base.
• PRINT LAST RECORD: Prints the last recorded sample using
the current configuration to determine the type of data printed (i.e.
cumulative vs. differential, raw vs. normalized, ft
3
vs. m
3
).
The print configuration is set through the PRINT SETUP button in
the Configuration screen. For more details about printing, see the
Configuration section later in this chapter.
• Date/Time: Displays the current date and time.
Note:
If the “X” appears
in the battery indicator
while the instrument is
counting, the pump will
stop automatically to
prevent the battery from
discharging completely.
• Battery Indicator: Indicates that the instrument is being powered
by a rechargeable battery. The amount of battery life left is denoted
by the fill inside the battery icon. When the battery is low, the
words “BATT LOW!” will appear on the screen and the unit will
beep continually until it is plugged into its AC power cord for
recharging. See Figure 4-4.
Figure 4-4 Battery Indicator Levels
• AC Indicator: Indicates that the instrument is being powered with
its AC power cord. If the battery is installed, the HANDHELD will
charge the battery when the instrument is plugged in.
• Flow Status: When the instrument is sampling, the Flow Indicator
will display sufficient or insufficient flow.
WARNING: If the Air Flow is insufficient, turn the instrument off and