The UCL calculation will be printed for each channel size if the
number of locations in the range is 2 through 9. The calculation will
only print if the Format settings (Setting screen) are set to CUML and
NORM. Any other format settings or locations less than 2 and greater
than 9 will print N/A for each channel.
View the same data in different formats just by changing the Particle
Display options on the Sample Setting screen.
The Data Summary can be viewed as:
Raw Counts, Differential Data
Raw Counts, Cumulative Data
Normalized Counts, Differential Data, Particles/ft
3
Normalized Counts, Differential Data, Particles/m
3
Normalized Counts, Cumulative Data, Particles/ft
3
Normalized Counts, Cumulative Data, Particles/m
3
PRINT RECORD
When you push the PRINT RECORD button, whatever record is
currently displayed in the Data screen will print as a single sample
record. The record will print according to the settings in the Print
Configuration. See Figure 4-62.
Figure 4-62 Examples of Output from Auto Printout of Sample
**HANDHELD**
Serial #: 30804001
-----------------------
Location: 29
08-02-03, 15:34:46
Sample Time: 00:01:00
Flow: 0.1 cfm
Laser: OK
74.7 DegF 52.2 %RH
Particle Counts:
Size Diff Cumul
0.3 29118 32543
0.5 2572 3425
1.0 797 853
3.0 29 56
5.0 24 27
10 0 3 3
**HANDHELD**
Serial #: 30804001
-----------------------
Location: 29
08-02-03, 15:34:46
Sample Time: 00:01:00
Flow: 0.1 cfm
Laser: OK
74.7 DegF 52.2 %RH
Particles/ Ft^3:
Size Diff Cumul
0.3 2222.0 2350.0
0.5 115.0 128.0
1.0 13.0 13.0
3.0 0.0 0.0
5.0 0.0 0.0
10 0 0 0 0 0